Datasheets |
Selected specifications |
 |
WSI-Corp
Porometer 3G and 3Gz |
- Methodology
- Pore Size: Liquid Displacement Porometer
using continually increasing pressure of dry gas to expel
wetting fluid.
- Permeability: Single pressure stabilization
for fixed time based on percent pressure or flow.
-
Pore size range:
-
3G: 500 - 0.05µm
-
3Gz: 500 - 0.04µm
-
Pressure ranges
|
 |
WSI-Corp Porometer Micro |
- Methodology
- Pore Size: Liquid Displacement Porometer
using continually increasing pressure of dry gas to expel
wetting fluid.
- Permeability: Single pressure stabilization
for fixed time based on percent pressure or flow.
- Pore Size Range: 100µm - 0.09µm diameter
- Pressure ranges: 0 - 0.3 bar and 0 - 7 bar
|
|
|
WSI-Corp Porometer Nano
(Please call or
e-mail for draft product information.) |
- Liquid-Liquid Displacement Porometer
- Pore Size Range: To 2 nm
The Nano is a product that focuses solely on the nanopore size range.
The limit for using gas to displace the liquid is about 18 nm because of
the high pressures required. Use of a liquid to displace the liquid
changes things a lot since the interfacial tension between the liquids
is a lot lower than the surface tension between a gas and liquid. This
means that for a given pressure the liquid liquid method measures
smaller pores. The <2 nm size is only applicable to this method and this
is only offered on the Nano. |
 |
Porometry Application Notes |